Loading…
Friday January 31, 2025 12:15pm - 2:15pm IST

Authors - Sudha K L, Navya Holla K, Kavita Guddad
Abstract - The antenna is a vital component of the Magnetic Resonance Imaging (MRI) machine which receives the radio frequency signals emitted by the protons in the body after the RF pulse is turned off. Specialized high frequency antennas can improve the quality, clarity, and resolution of the resulting MRI images. This paper deals with the design of Bow Tie antenna for X-Band in the frequency range 8–12 GHz, used in ultra-high field MRI systems. Using the Ansys HFSS tool, the antenna is designed and simulated and analysed. The fabricated antenna with the design specifications is tested in anechoic chamber for its working. Reflection coefficient at 10.5GHz is found to be around -14 dB for simulated antenna and -12 dB for fabricated antenna, which is satisfactory for practical application. Differences between the measured and simulated values were seen in results which are caused by cable loss in the measuring apparatus.
Paper Presenter
Friday January 31, 2025 12:15pm - 2:15pm IST
Virtual Room E Pune, India

Sign up or log in to save this to your schedule, view media, leave feedback and see who's attending!